Page 47 of 87 DGH 55B-INS-OMENG Rev: 3
13.2 Parameters for Continuous Averaging Measurement Mode
These are the Continuous Averaging Measurement Mode parameters that are accessible from the
Configuration Menu. The device must be set to Continuous Averaging to access these
parameters.
NOTE: To access these parameters, enter the Configuration Menu by pressing the CFG key.
Press the ENT key repeatedly to navigate through the menu. See section 13.1
13.2.1 Auto Switch (default enabled) can be enabled or
disabled for Continuous Averaging Mode. With
Auto Switch enabled, the device will automatically
switch eyes (after a 4 second delay) when a
measurement group is complete. Press the or
key to change the configuration.
13.2.2 Numb of Meas (default 25) is the number of
measurements that the unit will use to calculate the
measurement average and standard deviation of a
measurement point in Continuous Averaging
Measurement Mode. It can be adjusted from 1 to 25.
Press the or key to change the value.
WARNING: All measurements in the device, including those in memory will be cleared
when the Number of Measurements is changed.
13.2.3 Auto Rep Delay (default <50 msec) is the period of
time (in milliseconds) that the device will wait
between consecutive measurements while the probe
is applanated to the cornea in Continuous Averaging
Measurement Mode. It can be adjusted from
<50 msec to 950msec. Press the or key to
change the value.