Page 49 of 87 DGH 55B-INS-OMENG Rev: 3
13.3.5 Poor Appl Delay (default 2.0 sec) is the amount of
time that the device will wait before advancing to
the next mapping point after a poor applanation. It
can be adjusted from 1.0 sec to 9.5 sec. Press the
or key to change the value.
13.4 General Device Parameters
These are the general device parameters that are accessible from the Configuration Menu. They
are accessible at any time.
NOTE: To access these parameters, enter the Configuration Menu by pressing the CFG key.
Press the ENT key repeatedly to navigate through the menu. See section 13.1
13.4.1 Delay To Standby (default 1.0 min) is the amount
of time that the device will wait before going into
Standby Mode. It can be adjusted from 0.5 min to
9.5 min. Press the or key to change the value.
13.4.2 Battery Type should be selected to match the
batteries installed in the device. If the device detects
that the batteries have been removed, a prompt will
require the user to set the battery type accordingly
(see section 16.1.3). Press the or key to change
the value.
13.4.3 Contrast (default 7 bars) determines the contrast of
the screen. Press the or key to change the
value.
13.4.4 Volume (default 7 bars) determines the volume of
audible signals/alarms. Press the or key to
change the value.