4.1.8. WBC chamber draining process
The WBC chamber draining is performed via V3 (On) and V4 (On) valves by the pump while
the V5 must be off. The WBC chamber draining is always executed without using the puffer
reservoir. So other tasks like RBC chamber draining can be performed parallel with this
operation.