7.1. Error codes
%s: filename concerned, %d: error type concerned, %u: error specific string
7.1.1. Software/system errors
Error %d on opening file
%s
Fatal system or
program error!
The same as in case of error
1000.
Fatal system or
program error!
The same as in case of error
1000.
Data file %s is corrupt,
new will be created, all
stored measure data will
be lost
Stored data file is
corrupt or missing!
Disk or software
error.
New stored data file will be
created automatically (by User
confirmation).
Error %d on indexing file
%s
Fatal system or
program error!
The same as in case of error
1000.
Fatal pressure error.
Power off the system!
Cannot make the
measuring vacuum!!
Check the pump and the tubes,
fittings and valves around the
puffer reservoir.
Waste is full!
Empty waste container!
The system assumes that
you will do it before going
on!
Diluent reagent con-
tainer become empty,
or one of the diluent
fluid sensor is too
sensitive or there are
a lot of bubbles in the
diluent tubes!
Replace the Diluent reagent
container with a filled one.
If this error still remains with a
filled diluent reagent container,
calibrate sensors, and check the
tubings and the fittings.
Lyse reagent
containner become
empty, or the lyse
fluid sensor is too
sensitive or there are
lot of bubbles in the
lyse tubes!
Replace the Lyse reagent
container with a filled one.
If this error still remains with a
filled lyse reagent container,
calibrate sensors and check the
tubings and the fittings.