7.2.5. Hardware relating warnings
Code
Message
Reason(s)
Remedy
5400
+12V too low!
The +12V is not
sufficient, the
measurements are
unreliable.
Contact the Service Personnel.
5401
-12V too low!
The -12V is not
sufficient, the
measurements are
unreliable.
Contact the Service Personnel.
5402
Battery voltage too low!
The battery has been
discharged.
Check Date and Time settings.
This problem has no effect on
measurement.
Contact Service Personnel.
7.2.6. Other warnings
Code
Message
Reason(s)
Remedy
5901
Insert software update
media!
The system will restart.
Software upgrade
(USB).
After pressing the system
will restart.
5904
This function will delete
all selected records.
Delete.
If you press all records will
be deleted.
If you press it will be not
deleted.
5906
Now it is safe to turn the

Shut down process
has been finished.
Power off the analyzer
5907
Now it is time to wipe the
washing head with a
damp cloth. Make sure it
is clean, then you
proceed.
Maintenance day
You have to perform weekly
maintenance.
If you want to change the
maintenance day contact Service
Personnel.