7.1.5. Other errors
Code
Message
Reason(s)
Remedy
1900
You have to add at least
one user to use multi
user mode!
Multi-user problem
Follow the instruction, or change
to single user mode in the User
settings submenu.
1921
%s sensor calibration
failed!
The system turned off the
%s sensor.
Reagent container
became empty, or
there are lot of
bubbles in reagent
tubes.
Replace the reagent container
with a filled one.
If this is still a problem with a
filled reagent container, turn on
the sensor and try recalibrating
and check the tubings and the
fittings.