7.2. Warning messages
7.2.1. Pneumatic warnings
Code
Message
Reason(s)
Remedy
5001
Remove reagent tubing
at rear reagent inputs
(Diluent, Lyse, and
Cleaner).
First message of
preparing for
shipment.
Remove the reagent tubing and
leave waste connected.
After pressing the analyzer
drains itself.
5002
Connect min. 100 ml
distilled water to reagent
inputs using cleaning
tube kit.
Second message of
preparing for
shipment.
Connect cleaning tube kit with at
least 100 ml distilled water and
press . After this the
analyzer will be rinsed with the
distilled water.
5003
Remove cleaning tube
kit. Keep reagent inputs
free.
Third message of
preparing for
shipment.
Remove the reagent tubing and
leave waste connected.
After pressing the analyzer
drains itself.
5004
You can power off the
system!
The preparing for
shipment function is
finished.
Power off the instrument.
5005
Apertures are partially
clogged. Try cleaning!
During the
measurement one or
both of the apertures


aperture).
Try cleaning, perform blank
measurement. If the problem still
persists use the hard cleaning
function.
5006
Please empty waste
container!
The calculated waste
volume has reached
its maximum level.
Check and empty the waste
container. Press .
5008
There are no accepted
blank values.
Measure blank?
The last blank
measurement was
unsuccessful.
If you press you will exit
from measurement.
If you press the blank
measurement will start.
5009
The blank results are
expired!
Measure blank?
The reblank time has
expired.
If you press you will exit
from measurement.
If you press the blank
measurement will start.