EasyManua.ls Logo

Diatron Abacus + - Page 53

Default Icon
97 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Abacus
+
Service Manual
rev 1.09
53
Stretch histogram: setting this option to Yes
will modify the WBC histogram range from 0-
300 fl instead of 0-400 fl (this is recommended
for Veterinary mode).
Reagent code: This code can be used to
modify the offset of the HGB function. Enter a
value between 0 and 30 (default is 9).
Extended Probe Voltages: This option is ON by default. It makes the instrument accept the
physically highest probe voltages to allow operation at lower operating temperature range.
Maintenance day: You can select from week days. According to this setting, the instrument
will ask the user to do weekly maintenance during shut down.
Laboratory header lines: you can choose how many lines to contain the header of the
blood result.
Reagent pack: Using packed reagent in Reagent status menu system displays the reagent
pack installation day and the Reagent open stability period. User can change reagents only
is ignored.
Waste container capacity: it essential to set
the correct volume of the waste container for
proper usage of the softw,
code: 5006. Set this value two liters less than
the total volume of waste container.
Lyse container capacity: Enter here volume of lyse container.
Cleaner container capacity: Enter here volume of cleaner container.
Rinse container capacity: For future development. Enter here 0.
It is possible, but not recommended ignoring the reagent volume check entering 0 as volume
of containers.
Clogging detections, bubble volumes for
development, do not touch them!

Table of Contents

Related product manuals