DATA package
This package contains sample and patient data, the measured parameters and the markers
of histograms. One data element consists of two items: name and value, which are
separated by a <HT> character, and closed by <LF>:
SNO<HT>152<LF> internal identifier
DATE<HT>19980715<LF> date of measurement (YYYYMMDD)
TIME<HT>114500<LF> time of measurement (HHMMSS)
PID<HT>2<LF> patient identifier (0..9999)
NAME<HT>JOE SMITH<LF> patient name (max. 32 characters)
MODE<HT>0<LF> patient type
WRN<HT>0<LF> warning bits
1
(32-bit hexadecimal)
PM1<HT>12<LF> PLT lower marker
2
PM2<HT>204<LF> PLT upper marker
2
RM1<HT>51<LF> RBC lower marker
2
WM1<HT>23<LF> WBC upper marker
2
WM2<HT>57<LF> WBC LYM-MID marker
2
WM3<HT>92<LF> WBC MID-GRA marker
2
PARN<HT>22<LF> number of parameters to send
P01<HT> 6.6<HT>0<LF> first parameter ID
3
, value and flag
4
P02<HT>4.29<HT>0<LF> second parameter ID, value and flag
...
P22<HT> 8.2<HT>0<LF> last parameter ID, value and flag
Remarks:
1
Position and meaning of the warning flags:
Bit
Mask
Letter
Meaning
0
0x00001
c, q
RBC/PLT clogging.
3
0x00008
m
RBC/PLT Coincidence is too high. Linearity error.
6
0x00040
b
RBC Blank is high, or no RBC blank.
7
0x00080
p
PLT blank is high, or no PLT blank.
8
0x00100
C, Q
WBC clogging.
11
0x00800
M, N
WBC coincidence is too high. Linearity error.
14
0x04000
B
WBC Blank is high, or no WBC blank.
15
0x08000
H
HGB Blank is high, or no HGB blank.
16
0x10000
E
No WBC three part.
