Diatron MI PLC
rev 1.09
92
Data map of TmeasData inside an F package
Field name
explanation
note
no. of bytes
data type
RecNo
Internal Record ID
1
2
WORD
TimeStamp
Timestamp of sample
2
4
DateTime
OperatorID
Operator ID. Important in multiuser mode only.
1
2
WORD
SampleID
Sample ID
3
8
char
PatientID
Patient ID
3
20
char
Name
Patient name
3
32
char
BirthDate
Birth date of patient
2
4
DateTime
Sex
Gender of patient
4
1
BYTE
Doctor
Doctor's name
3
16
char
LimitsLow
Array of Low limits for parameters, 23 values
5
23 x 4 bytes
float array
LimitsHigh
Array of High limits for parameters, 23 values
5
23 x 4 bytes
float array
PatTypeName
Species name
3
20
char
VetMode
Internal ID of species
6
2
WORD
HistoRBC
Array of RBC histogram, 256 values
7
256
BYTE
HistoPLT
Array of PLT histogram, 256 values
7
256
BYTE
HistoWBC
Array of WBC histogram, 256 values
7
256
BYTE
PLTlo
Lower threshold of PLT graph
7
1
BYTE
PLThi
Upper threshold of PLT graph
7
1
BYTE
RBClo
Lower threshold of RBC graph
7
1
BYTE
WBClo
Lower threshold of WBC graph
7
1
BYTE
LYMhi
Upper threshold of LYMs on WBC graph
7
1
BYTE
GRAlo
Lower threshold of GRAs on WBC graph
7
1
BYTE
Param
Array of measured values (23 elements)
5
23 x 4 bytes
float array
Flag
Array of flags of measured values (23 elements)
8
23 x 1 bytes
BYTE array
Warning
Binary representation of measurement flags
9
8
DWORD
Lyse
Volume of lyse reagent used for sample, ml
5
4
float
Lyse_2
Volume of lyse reagent used for sample, ml
5
4
float
Options
Options set for sample
10
2
WORD
PrVMinW
Probe voltage min for WBC
11
2
WORD
PrVMaxW
Probe voltage max for WBC
11
2
WORD
PrVMinR
Probe voltage min for RBC
11
2
WORD
PrVMaxR
Probe voltage max for RBC
11
2
WORD
PrVMinW2
Probe voltage min for WBC2
11
2
WORD
PrVMaxW2
Probe voltage max for WBC2
11
2
WORD
Age
Age of patient as filled in PatID dialog
12
1
BYTE
ClogReport
Clogging information of sample
3
20
char