EasyManua.ls Logo

Dionex UltiMate 3000 - Preconcentration

Dionex UltiMate 3000
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
UltiMate 3000 RSLCnano Standard Applications
27
2.7.3 Installation Tips
Follow the General Recommendations for Applications ( page 6).
Using a stainless steel loop (on a standard system) instead of a PEEK loop is
recommended in case of injecting highly diluted samples.
With column oven temperatures below 45°C, TFA can be used instead of HFBA in the
loading solvent to load the sample onto the trap column.
Due to the speed of the separation, the value for the detector's time constant (response)
must be reduced to 0.1s.
Concentration of HFBA in the loading solvent can be increased to 0.1% in case trapping
problems are observed.
The loss of hydrophilic peptides (for example, peaks 1 and 2 are lower than expected)
may (partially) be decreased by lowering the oven temperature.
2.7.4 Testing the Application
The pre-concentration setup can be tested using the following conditions:
Property Setting
Mobile phase A 100% water + 0.05% TFA
Mobile phase B 50%/50% water/ACN + 0.04% TFA
Loading solvent 100% water + 0.05% HFBA
Sample Cytochrome C digest 1 pmol/μL, prepared according to the instruction sheet
Note: The sample must be diluted in the loading solvent.
Injection volume 0.5 μL (partial loop)
UV detection 214 nm
Loading time 3 min (may vary with different injection volume/routine)
Gradient 1% to 70% B in 8 min
90% B for 2 min
8.5 min equilibration
Oven temperature
60°C
WPS temperature
5°C (WPS-3000(B)T only)
Loading flow 10 μL/min
Flow rate 3.0 μL/min (capillary flow selector)

Table of Contents

Other manuals for Dionex UltiMate 3000

Related product manuals