102 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Operation
7.4.3 CBM mode
The CBM mode allows patient monitoring without unnecessary alarms during
extracorporeal oxygenation of the patient by a heart-lung machine.
Properties of CBM mode:
– All gas concentrations are measured independently of the respiratory phases.
–The CO
2 apnea alarms and pressure apnea alarms are inactive.
The CBM mode can be used in all active ventilation modes.
When ventilation modes are changed, the CBM mode remains active. Changing to
the standby mode deactivates the CBM mode.
Deactivating the CBM mode activates the apnea monitoring.
Activating
1. Open the Alarms dialog.
2. Touch the Settings tab (1).
3. For Cardiac bypass mode (CBM), touch the On button (3).
Deactivating
Deactivate the CBM mode optionally as follows:
► For Cardiac bypass mode (CBM), touch the Off button (2):
► In the main menu bar, touch the Exit CBM button.
35307
2 3
1