242 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
On the bottom side of the lower part of the breathing system:
Turn the black fresh-gas decoupling valve (3) counterclockwise and remove it.
13.9.2.6 List of disassembled components
After complete disassembly of the breathing system, the following machine-
reprocessable components are present:
Upper part of the breathing system
Lower part of the breathing system
Sensor cap of the O
2 sensor
–APL valve
Inspiratory port
Expiratory port
Inspiratory valve (yellow)
Expiratory valve (blue)
Fresh-gas decoupling valve (black)
Piston diaphragm
Breathing bag arm (option)
Optionally present are:
Absorber container and absorber insert
CLIC adapter (option)
13.10 Machine reprocessing
13.10.1 Preparation
Prerequisite: The breathing system and components are completely disassembled.
Information on the number of reprocessing cycles
At least 250 reprocessing cycles are possible for the breathing system.
37060

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals