242 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
On the bottom side of the lower part of the breathing system:
Turn the black fresh-gas decoupling valve (3) counterclockwise and remove it.
13.9.2.6 List of disassembled components
After complete disassembly of the breathing system, the following machine-
reprocessable components are present:
Upper part of the breathing system
Lower part of the breathing system
Sensor cap of the O
2 sensor
–APL valve
Inspiratory port
Expiratory port
Inspiratory valve (yellow)
Expiratory valve (blue)
Fresh-gas decoupling valve (black)
Piston diaphragm
Breathing bag arm (option)
Optionally present are:
Absorber container and absorber insert
CLIC adapter (option)
13.10 Machine reprocessing
13.10.1 Preparation
Prerequisite: The breathing system and components are completely disassembled.
Information on the number of reprocessing cycles
At least 250 reprocessing cycles are possible for the breathing system.