312 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Principles of operation
Changing the body weight during mechanical ventilation has no effect on the
current ventilation settings.
17.6.3 Influence of patient age
During operation, the set age influences the following:
– Calculation of MAC value
17.7 Support of Infinity ID accessories
The following accessories can be used if the device has the appropriate options:
– Infinity ID breathing circuit
– Infinity ID WaterLock 2 water trap
– Infinity ID flow sensors
– Infinity ID CLIC absorber
The Infinity ID functionality can be configured, see page 180.
17.7.1 Infinity ID functionality
17.7.1.1 Replacement interval monitoring
Automatic monitoring of the period of use is available for Infinity ID products. An
exceeded period of use is signaled during the system test.
The replacement interval for the connected Infinity ID accessories can be adjusted.
This interval must be specified in accordance with the applicable infection
prevention regulations or the requirements stated in the instructions for use for the
corresponding accessory.
17.7.1.2 Anti-interchange security
When Infinity ID breathing hoses and Infinity ID breathing bags are used, the
incorrect connection of breathing hoses and the breathing bag is detected and
reported. Hoses that are incorrectly connected with the breathing system trigger an
automatic alarm.