Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 225
Reprocessing
13.4 Reprocessing list
The following summary provides an overview of the reprocessing procedures
defined for each device component. For further information on the defined
reprocessing procedures, see chapters "Surface disinfection with cleaning" and
"Machine cleaning with thermal disinfection".
13.5 Reprocessing procedures
13.5.1 Validated reprocessing procedures
At the time of product-specific validation, the following reprocessing procedures
showed good material compatibility and effectiveness:
The effectiveness of the listed reprocessing procedures has been validated by
independent laboratories that certified to the standard ISO 17025.
Components Surface disinfection
with cleaning
Machine cleaning with
thermal disinfection
Device surfaces Yes No
Breathing system mount
1)
1) Special reprocessing measure, see page 236.
Yes No
Ports No Yes
Non-return valves (yellow,
blue, and black)
No Yes
APL valve Yes (exterior) Yes
Upper part of the breathing
system and lower part of the
breathing system
No Yes
Sensor cap of the O
2 sensor
2)
2) Special reprocessing measures, see page 245.
No Yes
Absorber container and
absorber insert
No Yes
Piston diaphragm No Yes
Breathing bag arm (option) Yes Yes
CLIC adapter (option) Yes Yes
Procedure Agent Manufac-
turer
Con-
centra-
tion
Con-
tact
time
Tem-
pera-
ture
Surface disinfection
with cleaning
Dismozon
plus
BODE Che-
mie
1.6 % 15 min –
Mikrobac
1)
1) For temperature measurement nozzles and pneumatic connections in the breathing system
mount (see "Disinfecting the breathing system mount", page 236)
BODE Che-
mie
–1 min–
Oxycide Ecolab USA 2.34 % 5 min –
Machine cleaning Neodisher
Mediclean
forte
Dr. Weigert 0.3 % 10 min 55 °C
(131 °F)
Machine disinfection
(thermal)
–––5min90°C
(194 °F)