Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 75
Assembly and preparation
After using the non-rebreathing system
1. Dismantle the non-rebreathing system and the sample line.
2. CAUTION
Risk due to leakage from an open AGS inlet nozzle
To prevent contamination of the ambient air with anesthetic gases, press the
sealing plug back into the inlet nozzle after using a non-rebreathing system.
Seal the inlet nozzle (4) on the AGS again with a sealing plug.
4.9 Connecting and replacing consumables
4.9.1 Disposable CO2 absorber
Connect or replace the CLIC absorber in accordance with its instructions for
use.
4.9.2 Reusable CO2 absorber
As an alternative to disposable CO2 absorbers, a reusable CO2 absorber may also
be used.
4.9.2.1 Dismounting and emptying
1. Unscrew the CO
2 absorber (1) from the breathing system.
2. Remove and dispose of the disposable dust filter (2), if present.
3. CAUTION
Risk of chemical burns
Soda lime is caustic and is a strong irritant for eyes, skin, and airway.
Handle the soda lime carefully and do not spill it.
Empty out the used soda lime and dispose of it according to its instructions for
use.
37107