268 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
16.2 General information
Units of measurement for pressure 1hPa = 1mbar = 1cmH2O
100kPa = 0.1MPa = 1bar = 1kPa x 100
User's operating location At the front at a distance of 1 m (39 in) and a
height of 1.5 m (59 in)
Tolerances All specified tolerances apply for ambient condi-
tions of 20 °C (68 °F), 60 % relative humidity,
and 1013 hPa (760 mmHg).
Accuracies The accuracies indicated below change accord-
ing to ambient pressure, temperature, and rela-
tive humidity. If one of the ambient conditions is
changed up to the permissible limit, the accu-
racy can change by up to 50 %. If more than
one of the ambient conditions is changed, the
accuracy may change by up to 100 %.
Example: Accuracy of a measured pressure
value: ±4 % under standard conditions. At
10 °C, the accuracy changes to ±6 %; at 10 °C
and 20 % relative humidity, to ±8 %.
Standardization All patient-related volumes and flows are based
on dry oxygen, converted to the conditions in
the lungs (BTPS).
16.3 Ambient conditions
During operation
Temperature 10 to 40 °C
(50 to 104 °F)
Ambient pressure 650 to 1060 hPa
(9.5 to 15.3 psi)
Relative humidity 20 to 95 %, non-condensing
CO
2 concentration 300 to 1000 ppm
Height above sea level Up to 3500 m (11483 ft)
During storage and transport
Temperature
Device without battery –20 °C to 60 °C
(–4°F to 140°F)
Battery –15 °C to 40 °C
(5 °F to 104 °F)
For storage longer than 12 months –15 °C to 25 °C
(5 °F to 77 °F)
Maximum storage duration without
recharging
180 days