182 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
Prerequisite: USB mass storage device has been connected to the USB port.
Open the System setup > Im/Export config. > Im/Export config. page (1).
The configurations saved on the USB mass storage device are displayed in a
list (4). If not all of the configurations are displayed, delete all configurations from
the USB mass storage device that are not needed or move them to a subdirectory
on the USB mass storage device.
The following settings are neither imported nor exported:
Device name
Date and time
IP address
11.4.1 Importing the configuration
1. Touch one of the configurations in the list (4).
2. Touch the Import button (2) and confirm.
3. Restart the device.
4. Check the device configuration for correctness.
11.4.2 Exporting the configuration
To export configurations, touch the Export button (3) and confirm.
11.5 Activating software options
The following software options require an activation code to be entered, followed by
activation:
Spontaneous breathing support
AutoFlow
Advanced trends
Advanced ventilation monitoring
Advanced gas monitoring
36030
4
1
3
2
1
System setup

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals