274 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Adult ventilation (typical VT >200 mL) Maximum compliance: 6.0 mL/hPa
(mL/cmH
2O)
Maximum length 200 cm (78.7 in)
350 cm (137.8 in) (leads to limitations in compli-
ance compensation and pressure accuracy)
Recommendation for a breathing bag hose
(if no breathing bag arm is being used)
Maximum length 180 cm (70.9 in)
350 cm (137.8 in) with increased fresh-gas con-
sumption
Recommendation for the breathing bag size
Volume 0.5 L to 5.0 L (at least double the value of the
tidal volume)
16.9 External fresh-gas outlet
Connection 22 mm (outer taper), 15 mm (inner taper), ISO
Delivery See "Fresh-gas delivery"
Pressure limitation Not pressure-limited
16.10 Anesthetic gas receiving system (AGS)
Information The device is equipped with an active or a pas-
sive anesthetic gas receiving system.
16.11 Active AGS
Information This system is designed for connection to an
anesthetic gas scavenging system which works
with a suction flow.
Suction flow
Normal range 30 to 50 L/min
At lower end of restricted range ≥10 L/min
Maximum fresh-gas flow to prevent contam-
inating ambient air
For external breathing systems (normal range) 9 L/min
For external breathing systems (restricted
range)
5L/min
16.8 Breathing system (continued)