EasyManuals Logo

Dräger Atlan A350 User Manual

Dräger Atlan A350
332 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #274 background imageLoading...
Page #274 background image
274 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Adult ventilation (typical VT >200 mL) Maximum compliance: 6.0 mL/hPa
(mL/cmH
2O)
Maximum length 200 cm (78.7 in)
350 cm (137.8 in) (leads to limitations in compli-
ance compensation and pressure accuracy)
Recommendation for a breathing bag hose
(if no breathing bag arm is being used)
Maximum length 180 cm (70.9 in)
350 cm (137.8 in) with increased fresh-gas con-
sumption
Recommendation for the breathing bag size
Volume 0.5 L to 5.0 L (at least double the value of the
tidal volume)
16.9 External fresh-gas outlet
Connection 22 mm (outer taper), 15 mm (inner taper), ISO
Delivery See "Fresh-gas delivery"
Pressure limitation Not pressure-limited
16.10 Anesthetic gas receiving system (AGS)
Information The device is equipped with an active or a pas-
sive anesthetic gas receiving system.
16.11 Active AGS
Information This system is designed for connection to an
anesthetic gas scavenging system which works
with a suction flow.
Suction flow
Normal range 30 to 50 L/min
At lower end of restricted range 10 L/min
Maximum fresh-gas flow to prevent contam-
inating ambient air
For external breathing systems (normal range) 9 L/min
For external breathing systems (restricted
range)
5L/min
16.8 Breathing system (continued)

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals