Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 281
Technical data
Accuracy ±(0.2 Vol% + 15 % relative)
±(0.2 kPa + 15 % relative)
Resolution of displayed value 0.1 Vol% (kPa) for desflurane
0.01 Vol% (kPa) for all other anesthetic gases
T
10...90 <500 ms
Anesthetic agent identification Automatic
Primary gas At the latest at 0.3 Vol%
Secondary gas At the latest at 0.4 Vol% or 0.1 xMAC (the
larger value applies)
With a desflurane concentration greater than
4 Vol%, mixture detection occurs at the latest
when the concentration of the second anes-
thetic gas rises above 10 % of the desflurane
concentration.
The secondary gas becomes the primary gas
when the expiratory xMAC value is more than
0.2 xMAC above that of the primary gas.
Minimum displayed concentration The specified detection thresholds refer to ris-
ing anesthetic gas concentrations (e.g., at the
start of surgery). If the anesthetic gas concen-
tration falls, a concentration of down to
0.05 Vol% will be measured, based on the last
anesthetic agent detected. Below this concen-
tration, a value of 0 Vol% will be displayed.
xMAC Based on the age of the patient, the anesthetic
gas concentration, and the nitrous oxide con-
centration (the xMAC value is corrected for
ambient pressure)
Range 0 to 9.9
Accuracy Refer to the accuracies of the respective mea-
sured gas values.
Resolution of displayed value 0.1
16.16 Gas measurement with O2 sensor
Information The oxygen measurement is performed in the
inspiratory limb of the breathing system and is
pressure-corrected.
Time after switch-on until the specified accu-
racy is attained
<180 s
Time after the insertion of the sensor cell until
the specified accuracy is attained
<30 min
Replacement interval of the oxygen cell 2 years
16.15 Gas measurement with PGM (continued)

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals