280 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Time until measured CO2 values are displayed
(with at least reduced accuracy)
<90 s
Typical delays <5 s
Cross sensitivity None with respect to alcohol (<3000 ppm in
blood),
Acetone (<1000 ppm), methane, water vapor,
NO and CO
Drift Compensated by automatic cyclic calibration
During the automatic calibration, ambient air is
fed to the breathing system and used as the
sample gas.
O
2
Range 0 to 100 Vol%
Accuracy ±(2.5 Vol% + 2.5 % relative)
Resolution of displayed value 1 Vol%
T
10...90 <500 ms
CO
2
Range 0 to 13.6 Vol%
0 to 13.6 kPa
0 to 102 mmHg
Accuracy ±(0.43 Vol% + 8 % relative)
±(0.43 kPa + 8 % relative)
±(3.3 mmHg + 8 % relative)
Resolution of displayed value 0.1 Vol%
0.1 kPa
1 mmHg
T
10...90 <350 ms
N
2O
Range 0 to 100 Vol%
Accuracy ±(2 Vol% + 8 % relative)
Resolution of displayed value 1 Vol%
T
10...90 <500 ms
Anesthetic gases
Range
Halothane 0 to 8.5 Vol% (kPa)
Isoflurane 0 to 8.5 Vol% (kPa)
Enflurane 0 to 10 Vol% (kPa)
Sevoflurane 0 to 10 Vol% (kPa)
Desflurane 0 to 20 Vol% (kPa)
16.15 Gas measurement with PGM (continued)

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals