Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 287
Technical data
Weight of the large version
Nominal configuration consisting of mechani-
cally controlled gas mixer, plug-in connector for
3 vaporizers, breathing system, CLIC adapter
and CLIC absorber, breathing hoses, central
supply hoses (5 m (16.4 ft)), scavenging hose
(5 m (16.4 ft))
Approx. 160 kg (353 lbs)
Permissible total weight 330 kg (727 lbs)
Touchscreen color screen
Screen diagonal Approx. 39 cm (15.3 in)
Background illumination LED
Resolution 1280 x 768 pixels
RFID system
Operating frequency 13.56 MHz ± 50 ppm (wideband)
Transmitter power 42 dBμA/m (200 mW ± 1 dB)
Modulation ASK (amplitude shift keying)
Electromagnetic compatibility Tested in compliance with IEC 60601-1-2
Protection classes
Device I, in compliance with IEC 60601-1
Applied parts (connections for breathing hoses) TYPE BF
Degree of protection IP20 according to IEC 60529,
meets ISO 80601-2-13
Classification in compliance with Directive
93/42/EEC, Annex IX
II b
UMDNS code 10-134
Use of latex The device is made without natural rubber
latex.
16.19 Interfaces and ports
Serial ports COM 1 and COM 2
Information Only connect devices that meet the require-
ments of IEC 60950-1 for ungrounded SELV
circuits and the requirements of IEC 60601-1
(as of the 2nd edition) for touchable secondary
circuits with a maximum nominal voltage of
24 V DC.
Use COM2 to connect devices for external
patient-gas measurement with the MEDIBUS
communication protocol.
16.18 Operating characteristics (continued)