260 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
13.13 Brief instructions for
reprocessing
The following information provides a quick overview of the
individual steps during reprocessing. Follow the relevant
chapters for detailed information and safety instructions.
Disassembling the patient-specific accessories
1. Unscrew the sample line.
2. Remove the breathing bag together with the breathing
bag hose.
3. Disassemble the breathing circuit and the filters.
Disposing of or reprocessing patient-specific accesso-
ries
4. Dispose of the following disposable accessories:
Sample line
Breathing circuit
Breathing bag
Filter
5. Reprocess the following reusable accessories:
Breathing circuit
Breathing bag
Surface disinfection of the device surfaces
6. Disinfect the following surfaces on the device:
Screen
Control dials of the vaporizers
Rotary knob
Control elements on the gas mixing unit
Work surface
Handles
Handles on drawers
Absorber
CLIC adapter (option)
Circuit plug on bag elbow
Water trap on the integrated patient-gas measure-
ment module (PGM)
APL valve
Breathing system cover
O
2+ key
Flow control valve on external O
2 flowmeter (option)
Pressure reducers (on the rear of the device)
Preparation for disassembly
7. Switch off the device and all devices connected to it.
8. Disconnect all power plugs.
Disassembling the water trap
9. Pull out and dispose of the water trap.
Disassembling the CO
2 absorber
10. Reusable CO
2 absorber
a. Unscrew the absorber.
b. Remove and dispose of the disposable dust filter.
c. Empty out and dispose of the used soda lime.
d. Remove the absorber insert from the absorber con-
tainer.
11. CLIC absorber (disposable absorber)
a. Remove and dispose of the CLIC absorber.
Disassembling the breathing bag arm (option)
12. Unscrew the breathing bag arm (option).
Disassembling the O
2 sensor
13. Disassemble the O
2 sensor.
14. Remove the sensor cap.
Removing the breathing system
15. Remove the breathing system.
Removing the piston diaphragm
16. Remove the piston diaphragm.
Disinfecting the breathing system mount
17. Disinfect the breathing system mount.
Disassembling the breathing system
18. CLIC adapter (option)
a. Unscrew the adapter.
b. Insert the cleaning plate and close the holder.
19. Disassembling the flow sensors and ports
a. Remove the expiratory port.
b. Remove the expiratory flow sensor.
c. Remove the inspiratory port.
d. Remove the inspiratory flow sensor.
20. Lift off the upper part of the breathing system.
21. Screw off the APL valve.
22. Remove the non-return valves.
Inspiratory valve (yellow)
Expiratory valve (blue)
23. Remove the fresh-gas decoupling valve (black).
Machine reprocessing
24. Position the following parts appropriately and perform
machine reprocessing:
Upper part of the breathing system
Lower part of the breathing system
Sensor cap of the O
2 sensor
APL valve
Inspiratory port

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals