88 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Getting started
6.5 Emergency start-up
6.5.1 Electronically controlled gas mixer
1. Adjust the APL valve.
2. Set the O
2 switch (3) to the Add. O2 position.
3. Open the flow control valve (2) and set the desired O
2 flow. If necessary, press
the O
2+ key (4) to quickly fill the breathing bag.
4. Monitor the set flow on the O
2 flowmeter (1).
5. Set the anesthetic gas concentration on the vaporizer.
6. Ventilate the patient manually.
7. Turn on the device.
8. As soon as the Standby page is displayed, start the therapy, see page 93.
9. Set the O
2 switch (3) to Aux. O2.
10.Close the flow control valve (2).
6.5.2 Mechanically controlled gas mixer
1. Adjust the APL valve.
2. Open the flow control valve (1) and set the desired O
2 flow. If necessary, press
the O
2+ key (3) to quickly fill the breathing bag.
3. Monitor the flow on the total flow tube (2).
4. Set the anesthetic gas concentration on the vaporizer.
5. Ventilate the patient manually.
6. Turn on the device.
7. As soon as the Standby page is displayed, start the therapy, see page 93.
35041
3
2
1
4
35054
1
3
2

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals