EasyManuals Logo

Dräger Atlan A350 User Manual

Dräger Atlan A350
332 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #199 background imageLoading...
Page #199 background image
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 199
Troubleshooting
12.10 Problems with the active anesthetic gas receiving
system (AGS)
12.10.1 Replacing the particle filter of the active anesthetic gas receiving
system (AGS)
1. Fold down the cover (1).
2. Withdraw the particle filter (2).
3. Insert the cleaned or new particle filter.
Fault Cause Remedy
Flow indicator is below the
restricted range.
The suction power of the
ejector in the terminal unit
of the anesthetic gas
scavenging system
(AGSS) is insufficient.
Have the function of the
AGSS terminal unit
checked. Follow the corre-
sponding instructions for
use.
The particle filter is con-
taminated or blocked.
Replace the particle filter.
The flow indicator is above
the normal range.
The suction power of the
ejector in the AGSS termi-
nal unit is too high.
Have the suction power of
the ejector in the AGSS
terminal unit adjusted to
the working range of the
AGS.
The particle filter is miss-
ing.
If no soiling can be seen in
the transparent housing of
the AGS or on the flow
indicator, insert the parti-
cle filter.
If soiling can be seen in
the transparent housing of
the AGS or on the flow
indicator, replace the
AGS.
38127

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals