EasyManuals Logo

Dräger Atlan A350 User Manual

Dräger Atlan A350
332 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #200 background imageLoading...
Page #200 background image
200 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Troubleshooting
4. CAUTION
Risk of contaminating the ambient air
A filter cover that is not properly closed can cause contamination of the ambient
air with anesthetic gases.
Take care that the cover is properly closed.
Fold up the cover.
12.10.2 Replacing the anesthetic gas receiving system (AGS)
This chapter describes the replacement using the example of an active AGS which
is connected to the device with the transfer hose and to the anesthetic gas
scavenging system with the scavenging hose.
Disassembly
1. Disconnect the transfer hose (1).
2. Disconnect the scavenging hose (2).
3. Unfasten the 2 screws (3).
4. Withdraw the AGS to the side.
5. Dispose of the AGS.
The AGS must not be reprocessed.
Assembly
1. Slide in the new AGS.
2. Fasten the AGS with 2 screws.
3. Connect the scavenging hose.
4. Connect the transfer hose.
38134

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals