Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 159
Alarms
10.4 Adopting alarm settings when changing the ventilation
mode
When the ventilation mode is changed, the alarm settings are also adapted.
Depending on the mode, alarm settings can either be adopted or set to Off.
For some modes, it can be configured whether or not the settings are adopted.
Observe the information in the following section: "Vertical tab "Config. 2"", page 171
However, the settings can be adjusted during operation at any time.
Alarm or alarm limit CBM mode Pause Man/Spon,
PSV with Psupp
<5 hPa (cmH
2O)
Ext. FGO VC,
VC - AF,
PC,
PSV with Psupp 5hPa
(cmH
2O)
FiO
2 low
1)
1) With inspiratory O2 measurement with O2 sensor
Is adopted. Is adopted.
Not measured
Is restored or remains active.
FiO
2 high
1)
Configurable Configurable
FiO
2 low
2)
2) With integrated patient-gas measurement module
Is adopted.
inAgent high
2)
Apnea (no CO2)
2)
Off
Off
3)
3) The alarm remains suppressed until respiratory activity is detected.
Is adopted.
etCO
2 high
2)
Configurable
etCO
2 low
2)
inCO2 high
2)
FiO2 high
2)
Configurable
inAgent low
2)
xMAC low
2)
Not activated
Paw high
Is adopted. Is adopted.
Not measured
Paw low
MV high
Configurable
Off
Configurable
MV low
Apnea (no flow)
Configu-
rable
4)
4) This alarm is only activated when the MV low alarm limit is also activated.
Off3)
On
Apnea (no pressure) Off Off