Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 171
Configuration
11.3.3.5 Vertical tab "Config. 2"
11.3.3.6 Vertical tab "Config. 3"
Headline/
Parameter
Setting range
Description
Deactivate the alarm limit in the following modes: Man/Spon,
Pause, CPAP/PSV with Psupp < 5, ext. FGO
FiO
2 high Yes; No Yes; No Yes; No Specifies the alarm behavior when
changing to a different ventilation
mode.
These settings apply only to a change
to a ventilation mode with lower or no
breathing support (see page 299).
The alarm behavior at the start of the
therapy is defined by the configuration
in the vertical tab Limits.
MV low Yes; No
MV high Yes; No
xMAC low Yes; No
etCO
2 low Yes; No
etCO
2 high Yes; No
inCO
2 high Yes; No
inAgent low Yes; No
Deactivate the alarm limit in cardiac bypass mode (CBM)?
FiO
2 high Yes; No Specifies the alarm behavior in the
CBM mode.
MV low Yes; No
MV high Yes; No
inAgent low Yes; No
Headline/
Parameter
Setting range
Description
Alarm limits for "Cylinder almost empty"
O
2 [bar]; [kPax100]: Off; 15 to 50
20
[psi]: Off; 218 to 725
290
Specifies the alarm limits for the supply
pressure of connected gas cylinders.
Air [bar]; [kPax100]: Off; 15 to 50
20
[psi]: Off; 218 to 725
290
N
2O [bar]; [kPax100]: Off; 15 to 40
20
[psi]: Off; 218 to 580
290

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals