Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 67
Assembly and preparation
Connect the scavenging hose (1) to the nozzle on the AGS.
2. Secure the scavenging hose with the clip (2).
3. Lay the scavenging hose so that the gas is disposed of, e.g., by a ventilation
system.
4. Fasten the end of the hose.
4.7 Preparation for an operation day / after cleaning and
sterilization
Assemble the device and prepare it ready for operation.
Observe the information in the following chapter: "Fitting and
assembly", page 249.
4.8 Selecting and connecting patient-specific accessories
4.8.1 Fitting the breathing circuit and the filters
This device is made without natural rubber latex.
To minimize the risk of contact with latex, use breathing bags and breathing hoses
that are not made with natural rubber latex.
The device can be used with Infinity ID breathing hoses or conventional breathing
hoses. If no leakage test has yet been performed after switching on the device,
hose compliance and hose resistance will automatically be adopted when Infinity ID
breathing hoses are connected.
Do not use any inspiratory or expiratory bacteria filters if the ID functionality of
the Infinity ID breathing circuit is to be used. In this case, fit a filter to the Y-piece.
In cases which preclude use of a bacteria filter at the Y-piece, the Infinity ID function
of the Infinity ID breathing circuit cannot be used.