Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 55
Assembly and preparation
4.3 Before first operation
The breathing system must be reprocessed before the device is operated for the
first time.
4.3.1 Inserting the O2 sensor cell
If the device is equipped with an O2 sensor, an O2 sensor cell must be inserted into
the O
2 sensor.
1. Disassemble the O
2 sensor (see "Disassembling the O2 sensor", page 233).
2. Insert the new O
2 sensor cell (1) into the O2 sensor.
After the O2 sensor cell has been inserted, the initialization phase of the O2
sensor takes place. The initialization lasts 30 minutes. The 30 minutes start when
the mains power supply is established.
4.3.2 Assembling and inserting the O2 sensor
1. Place the sensor cap (1) on the O2 sensor.