Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 275
Technical data
For internal breathing systems (restricted
range)
7L/min
Connection for sample gas disposal when
using external patient-gas measurement
Outer hose diameter 3 to 6 mm
Maximum inlet flow 500 mL/min
16.12 Passive AGS
Information This system is designed for an anesthetic gas
scavenging system which works without suction
flow. Do not connect it to an active anesthetic
gas scavenging system.
To limit contamination of the ambient air in com-
pliance with ISO 80601-2-13, follow the specifi-
cations for the scavenging hose.
Scavenging hose
Maximum length 8 m (26 ft)
Minimum diameter 19 mm (0.75 in)
Connection for sample gas disposal when
using external patient-gas measurement
Outer hose diameter 3 to 6 mm
Maximum inlet flow 500 mL/min
16.13 Measuring systems and displays
Information Alarms and internal calculations are based on
values that are more precise than the displayed
values. Consequently, there may be minor devi-
ations between the current alarm status and the
displayed measured values.
Airway pressure
Airway pressure Paw
Plateau pressure Pplat
Positive end-expiratory pressure PEEP
Peak inspiratory pressure PIP
Mean airway pressure Pmean
Range –20 to +99 hPa (cmH
2O)
Accuracy ±4 % of the measured value or ±2 hPa
(cmH
2O) (the larger value applies)
16.11 Active AGS (continued)