Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 87
Getting started
For further information about the configuration, see page 166.
6.4 Checking the operational readiness
Using colors, the Standby page (1) indicates whether the system test was
successful and the device is ready for operation.
If the device is not fully operational, the most important irregularities (5) are
displayed along with a recommendation to perform a specific test (3).
Additionally, the current system leakage is displayed in area (5).
To view details regarding the status of the device, touch the Details... button (2) or
the Tests... button (4), see page 126.
Dräger recommends performing the system test every 24 hours. Otherwise, it will
not be possible to ensure that the device is functional.
35034
Color Meaning
Green System is fully operational.
Yellow System is operational with limitations.
There are functional restrictions.
Take further measures to ensure patient safety (e.g., external
monitoring).
Red System is not operational.
Contact service personnel if necessary.
2
1
3
4
5

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals