86 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Getting started
6 Getting started
6.1 Safety instructions
Checking the safety systems
Some safety systems are only checked during start-up. If this check is not
performed regularly, a device malfunction may occur, putting the patient at risk.
Restart the device at least once per month so that the safety systems will be
checked regularly.
6.2 Turning on the device
Prerequisite: The device has been reprocessed (see page 222) and assembled
ready for operation (see page 45).
1. Connect the device to the mains power supply.
2. Set the main switch to position I.
3. CAUTION
Risk of device malfunction
Condensed water may form when the device is brought from a cold storage
location into a warm environment.
To prevent condensation and resulting failures of electrical components, do
not turn on the device after abrupt temperature changes for 1 to 2 hours.
Press the key (1).
The device starts. The Standby page is displayed.
If there is sufficient battery charge, the device will also start without the power plug
being plugged in.
6.3 Checking the device configuration
The device can be customized to suit the requirements of the user. Settings for the
following features are possible:
Start settings for the ventilation
Alarm limits
General device behavior
35026
1

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals