Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 227
Reprocessing
Dräger states that oxygen-releasing agents and chlorine-releasing agents may
cause color change in some materials. Color change does not indicate that the
product is not functioning correctly.
13.6 Preparation
13.6.1 Disassembling the patient-specific accessories
1. WARNING
Risk of infection
Used sample lines and water traps may be infectious due to the breathing gases
that passed through them.
Replace the sample line and the water trap regularly in the following
situations:
If the sample line is connected to the filter on the Y-piece, replace it daily.
If there is no filter fitted to the Y-piece and the sample line is connected
directly to the Y-piece, replace the sample line after every patient.
Remove the sample line from the water trap.
Initially leave the water trap fitted to prevent infectious fluid from spurting out.
Remove the water trap only after surface disinfection.
Replace the water trap after each patient.
Quaternary ammonium
compounds
Mikrozid sensitive liquid
1)
Schülke & Mayr
Mikrozid sensitive wipes
1)
Mikrozid alcohol-free liquid
1)
Mikrozid alcohol-free wipes
1)
acryl-des
1)
Cleanisept Wipes Maxi Dr. Schumacher
Surfa'Safe Premium ANIOS Laboratories
Wip'Anios Excel
Tuffie 5 Vernacare
1) Virucidal against enveloped viruses
Class of active ingre-
dient
Surface disinfectants Manufacturer

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals