Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 227
Reprocessing
Dräger states that oxygen-releasing agents and chlorine-releasing agents may
cause color change in some materials. Color change does not indicate that the
product is not functioning correctly.
13.6 Preparation
13.6.1 Disassembling the patient-specific accessories
1. WARNING
Risk of infection
Used sample lines and water traps may be infectious due to the breathing gases
that passed through them.
► Replace the sample line and the water trap regularly in the following
situations:
– If the sample line is connected to the filter on the Y-piece, replace it daily.
– If there is no filter fitted to the Y-piece and the sample line is connected
directly to the Y-piece, replace the sample line after every patient.
► Remove the sample line from the water trap.
► Initially leave the water trap fitted to prevent infectious fluid from spurting out.
Remove the water trap only after surface disinfection.
► Replace the water trap after each patient.
Quaternary ammonium
compounds
Mikrozid sensitive liquid
1)
Schülke & Mayr
Mikrozid sensitive wipes
1)
Mikrozid alcohol-free liquid
1)
Mikrozid alcohol-free wipes
1)
acryl-des
1)
Cleanisept Wipes Maxi Dr. Schumacher
Surfa'Safe Premium ANIOS Laboratories
Wip'Anios Excel
Tuffie 5 Vernacare
1) Virucidal against enveloped viruses
Class of active ingre-
dient
Surface disinfectants Manufacturer