Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 229
Reprocessing
13.7 Surface disinfection with cleaning
13.7.1 Surface disinfection of the device surfaces
1. Remove soiling immediately. Use a cloth dampened with disinfectant to remove
soiling.
2. WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
Damage to the device
Electric shock
Device malfunctions
Ensure that no liquid penetrates the device.
CAUTION
Risk of contaminating the device
If the water trap is removed, fluid may get into the main device while it is being
disinfected.
Leave the water trap fitted during surface disinfection.
Perform the surface disinfection of the entire device by wiping backwards and
forwards at least 3 times:
In particular, disinfect device surfaces and hand contact points that are
touched frequently.
Follow the list of validated disinfectants (see "Validated reprocessing
procedures", page 225).
The following illustration shows the frequently touched surfaces and hand
contact points:
39472
No. Designation
1 Screen
2 Control dials of the vaporizers

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals