Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 289
Technical data
Distribution piece for central O2 supply
(optional)
Supply pressure 2.7 to 6.9 kPa x 100
39 to 100 psi
0.27 to 0.69 MPa
Maximum permissible flow 20 L/min
16.20 Relevant standards
Information In addition to the standards listed here, this
medical device also meets various other stan-
dards, e.g., standards concerning special
national requirements.
IEC 60601-1
Medical electrical equipment
Part 1:
General requirements for basic safety and
essential performance
IEC 60601-1-2
Medical electrical equipment
Part 1-2:
General requirements for basic safety and
essential performance -
Collateral standard: Electromagnetic
disturbances - Requirements and tests
IEC 60601-1-8
Medical electrical equipment
Part 1-8:
General requirements for basic safety and
essential performance - Collateral standard
Alarm systems - General requirements, tests
and guidance for alarm systems in medical
electrical equipment and medical electrical sys-
tems
IEC 60601-2-13
Medical electrical equipment
Part 2-13:
Particular requirements for the safety of anaes-
thetic systems
ISO 8835-2
Systems for inhalational anesthesia
Part 2:
Anesthesia breathing systems
ISO 8835-3
Systems for inhalational anesthesia
Part 3:
Transfer and receiving systems of active
anaesthetic gas scavenging systems
ISO 8835-4
Systems for inhalational anesthesia
Part 4:
Anesthetic vaporizers
ISO 8835-5
Systems for inhalational anesthesia
Part 5:
Anesthesia ventilators
ISO 21647
Medical electrical equipment
Particular requirements for the basic safety and
essential performance of respiratory gas moni-
tors
16.19 Interfaces and ports (continued)