282 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Cross sensitivity None with respect to alcohol (<3000 ppm in
blood),
Acetone (<1000 ppm), methane, water vapor,
NO and CO
Drift Compensated by cyclic calibration with ambient
air, at the latest every 7 days in the system test
O
2
Range 0 to 100 Vol%
Accuracy ±(2.5 Vol% + 2.5 % relative)
Resolution of displayed value 1 Vol%
Typical delays <15 s
16.17 Display of calculated values
Measurement of consumption and elimina-
tion
CO
2 elimination per minute
Standardization The following data are standardized to STPD
conditions.
Range 0 to 9999 mL/min
Accuracy ±25 % or ±100 mL/min (the larger value
applies)
Resolution of displayed value 1 mL/min
O
2 uptake per minute
Standardization The following data are standardized to STPD
conditions.
Range 0 to 9999 mL/min
Accuracy ±25 % or ±100 mL/min (the larger value
applies)
Resolution of displayed value 1 mL/min
Anesthetic agent uptake
Range 0 to 99.9 mL fluid
Accuracy ±25 % or ±1 mL (the larger value applies)
Resolution of displayed value 0.1 mL
Fresh-gas consumption
Standardization The following data are standardized to STPD
conditions.
Range 0 to 99999 L
16.16 Gas measurement with O2 sensor (continued)