Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 273
Technical data
In mechanical ventilation modes Typically 0.5 mL/hPa (mL/cmH2O)
corresponds to 15 mL at 30 hPa (cmH
2O)
for the "Neonates" patient category
Typically 0.8 mL/hPa (mL/cmH
2O)
corresponds to 24 mL at 30 hPa (cmH
2O)
for the "Pediatric patients" patient category
Typically 2.0 mL/hPa (mL/cmH
2O)
corresponds to 60 mL at 30 hPa (cmH
2O)
for the "Adults" patient category
Filling volume of the CO
2 absorber
Reusable CO
2 absorber 1500 mL ± 50 mL
Disposable CO
2 absorber CLIC absorber 800+ 1300 mL ± 50 mL
Disposable CO
2 absorber CLIC absorber Free 1200 mL ± 50 mL
Flexible breathing bag arm
Volume 0.11 L ± 50 mL
Compliance 0.11 ± 0.05 mL/hPa (mL/cmH
2O)
corresponds to 3.3 ± 1.5 mL at 30 hPa
(cmH
2O)
Rigid breathing bag arm
Volume 0.13 L ± 50 mL
Compliance 0.13 ± 0.05 mL/hPa (mL/cmH
2O)
corresponds to 3.9 ± 1.5 mL at 30 hPa
(cmH
2O)
Total leakage <150 mL/min at 30 hPa (cmH
2O) standardized
to BTPS conditions
APL valve
Setting range Open, 5 to 70 hPa (cmH
2O)
Accuracy (at a flow of 20 ±1 L/min): ±20 % of the set value or ±3 hPa (the larger
value applies), but not more than +10 hPa
(cmH
2O)
Pressure drop at 30 L/min (ATPD), fully opened Dry: 2.1 hPa (cmH
2O)
Wet: 2.2 hPa (cmH
2O)
Recommendation for breathing hoses
Information All compliances and volumes specified include
inspiratory and expiratory filters.
Neonate ventilation (typical VT <100 mL) Maximum compliance: 2.0 mL/hPa
(mL/cmH
2O)
Pediatric ventilation (typical VT between 50 and
300 mL)
Maximum compliance: 4.0 mL/hPa
(mL/cmH
2O)
16.8 Breathing system (continued)