236 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
Remove the breathing system (2) vertically upwards from the breathing system
mount.
13.8.7 Removing the piston diaphragm
Remove the piston diaphragm (1) from the ventilator. If necessary, remove any
residual moisture present.
13.8.8 Disinfecting the breathing system mount
Note the following information:
Do not use alcohol-based disinfectants.
Remove any residues.
Ensure lint-free cleaning and disinfection.
Do not allow fluid to penetrate.
Prerequisite:
The breathing system cover has been removed.
The breathing system has been removed.