180 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
11.3.6.7 Vertical tab "Infinity ID"
11.3.6.8 Vertical tab "Service"
11.3.7 Resetting the start settings
Certain pages in the System setup dialog have a button for resetting the respective
start settings to the factory defaults.
11.3.7.1 Resetting the changes on a page
1. Open the appropriate tab.
2. Touch the Factory defaults button and confirm.
11.3.8 Resetting the consumptions
The gas consumption can be reset in Standby > System setup > System status >
Consumption.
Touch the Reset data button and confirm.
Headline/
Parameter
Setting range
Description
Monitoring of Infinity ID accessories
Breathing circuit On; Off Activates or deactivates the Infinity ID
functionality.
On:
Generates a message when the
maximum period of use is exceeded
Generates a message when
Infinity ID breathing hoses are incor-
rectly connected
Off:
Messages are suppressed.
Water trap On; Off
Flow sensors On; Off
CO
2 absorber On; Off
Exchange interval [days]
Breathing circuit Off; 2 to 9
2
Specifies the replacement intervals for
Infinity ID accessories.
Water trap Off; 28
Flow sensors Off; 1 to 180
90
CO
2 absorber Off; 1 to 28
7
Headline/
Parameter
Setting range
Description
Service access code The following functions are available
after the appropriate credentials have
been entered:
Access to the service dialog

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals