Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 243
Reprocessing
After complete disassembly of the breathing system, the following machine-
reprocessable components are present:
Upper part of the breathing system
Lower part of the breathing system
Sensor cap of the O
2 sensor
–APL valve
Inspiratory port
Expiratory port
Inspiratory valve (yellow)
Expiratory valve (blue)
Fresh-gas decoupling valve (black)
Piston diaphragm
Breathing bag arm (option)
Optionally present are:
Absorber container and absorber insert
CLIC adapter (option)
Required aids:
Hose for connecting the breathing bag arm
Hose for connecting the absorber insert
13.10.2 Positioning the components
Depending on device configuration, it is possible that not all components will be
present.
Position the components securely in the load carrier. Ensure the following:
All surfaces and interior spaces can be flushed completely.
The water can drain off freely.
The following illustrations are nonbinding recommendations. Depending on the
washer-disinfector used, alternative positioning is possible.

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals