Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 117
Operation
Automatic setting
Alarm limits can be automatically adapted to current measured values or set values.
1. Open the Alarms > Limits dialog.
2. To adjust the alarm limits for one individual parameter, touch and confirm one of
the Autoset buttons (5).
To adjust the alarm limits for all parameters, touch and confirm the Autoset all
button (4).
The function for adjusting all alarm limits can be called up directly with the Autoset
limits button in the main menu bar.
Only use the automatic adjustment when measured values and set values are
stable.
The lower alarm limit for the xMAC level is also adjusted during automatic setting,
see page 118.
Configuration and algorithm, see page 169.
7.9.2 Activating and deactivating CO2 alarms
Prerequisite: The device has the "Integrated patient-gas measurement module"
option.
The CO
2 monitoring (alarms for inCO2, etCO2, and CO2 apnea) can be activated or
deactivated.
1. Open the Alarms dialog.
2. Touch the Settings tab (1).
3. For CO
2 alarms, touch the button (2):
On: Alarms are activated.
Off: Alarms are deactivated.
Or
Use the CO
2 alarms off button in the main menu bar to activate or deactivate
the alarms.
35936
1
2
Alarms

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals