Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 145
Tests
8.4.5.3 Manual resuscitator
1. Make sure there is a manual resuscitator on the device.
2. Check the functional integrity of the manual resuscitator.
8.4.5.4 Anesthetic gas receiving system
With active anesthetic gas scavenging:
Have the flow for the anesthetic gas scavenging system set so that the red flow
indicator (1) floats in range (2) ("normal range").
If the flow indicator (1) is floating in range (3) ("restricted range"), certain fresh-
gas flows should not be exceeded, see "Anesthetic gas receiving system" in
chapter "Technical data". Contamination of the ambient air can be prevented by
limiting the fresh-gas flow.