Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 147
Tests
Without Infinity ID function:
Absorbers will not be detected, e.g., reusable CO
2 absorbers.
Update the replacement date manually: Touch the Reset button after the
soda lime has been replaced.
Only perform leakage tests with the CLIC absorber locked into place because
the CLIC absorber affects the system compliance values.
8.4.6.2 Sample line
Check that the sample line is correctly connected.
8.4.6.3 Water trap
1. Check the water level in the water trap.
2. Check the period of use of the water trap. Replace the water trap when
necessary.
With Infinity ID function:
Water traps of type Infinity ID WaterLock 2 will automatically be detected and
the replacement date will automatically be set.
3512935130