168 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Configuration
11.3.3 System setup > Alarms
11.3.3.1 Vertical tab "Limits"
Headline/
Parameter
Setting range
Description
FiO
2 [%]
19 to 99;
Off
19 to 99;
Off
19 to 99;
Off
90
Inspiratory oxygen fraction
FiO
2 [%]
18 to 98
20
18 to 98
20
18 to 98
20
etCO
2
[%]; [kPa]
[mmHg]
0.1 to 9.8;
Off
7.0
1 to 75;
Off
53
0.1 to 9.8;
Off
7.0
1 to 75;
Off
53
0.1 to 9.8;
Off
7.0
1 to 75;
Off
53
Expiratory CO
2 concentration
etCO
2
[%]; [kPa]
[mmHg]
Off;
0.0 to 9.7;
Off;
0 to 74
Off;
0.0 to 9.7;
Off;
0 to 74
Off;
0.0 to 9.7;
Off;
0 to 74
inCO
2
[%]; [kPa]
[mmHg]
0.1 to 1.4;
Off
1.1
1 to 10;
Off
8
0.1 to 1.4;
Off
1.1
1 to 10;
Off
8
0.1 to 1.4;
Off
1.1
1 to 10;
Off
8
Inspiratory CO
2 concentration
Paw high
[mbar]; [hPa];
[cmH2O]
5 to 99
40
5 to 99
25
5 to 99
20
Airway pressure
If the airway pressure is above the
value set for Paw low for longer than
15 s, the Airway press. continuously
high alarm is issued.
Paw low
[mbar]; [hPa];
[cmH
2O]
Auto;
3 to 97
Auto;
3 to 97
Auto;
3 to 97
inSev
[%]; [kPa]
0.10 to 9.95
4.40
0.10 to 9.95
5.10
0.10 to 9.95
6.70
Sevoflurane
inSev
[%]; [kPa]
Off;
0.00 to 9.85
Off;
0.00 to 9.85
Off;
0.00 to 9.85
inDes
[%]; [kPa]
0.1 to 20.0
12.5
0.1 to 20.0
14.5
0.1 to 20.0
19.0
Desflurane
inDes
[%]; [kPa]
Off;
0.0 to 19.9
Off;
0.0 to 19.9
Off;
0.0 to 19.9

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals