246 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
Non-return valves and fresh-gas decoupling valve
► Position the valves with the valve cages facing downwards.
Breathing bag arm (option)
1. Connect the breathing bag arm (1) to a hose (2).
2. Lay the breathing bag arm together with the hose in the load carrier.
3. Connect the hose (2) to a flushing nozzle (3).
Absorber insert
1. Connect the absorber insert (1) to a hose (2).
2. Lay the absorber insert together with the hose in the load carrier.
3. Connect the hose (2) to a flushing nozzle (3).
395013950239503