248 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
Thermal disinfection
–Drying
Proceed as follows after the cycle has ended:
1. Check all components for visible soiling.
2. Repeat the cycle if necessary.
13.10.4 Drying the components
Allow all the components to dry completely.
To prevent residual moisture, adjust the drying times during machine
reprocessing.
13.10.5 Additional reprocessing procedures
Additional reprocessing procedures have not been tested by Dräger with regard to
their effectiveness.
However, the materials in the following components permit steam sterilization up to
a maximum of 137 °C (278.6 °F):
Upper part of the breathing system
Lower part of the breathing system
Sensor cap of the O
2 sensor
–APL valve
Inspiratory port
Expiratory port
Inspiratory valve (yellow)
Expiratory valve (blue)
Fresh-gas decoupling valve (black)
Piston diaphragm
Breathing bag arm (option)
Absorber container
Absorber insert
After sterilization, inspect the components for damage and replace if necessary.

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals