Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 303
Principles of operation
17.2.5 Pressure-controlled ventilation
17.2.5.1 PC
17.2.5.2 PC - CMV
Pressure-controlled
Time-controlled
Machine-triggered
The mandatory breaths are machine-triggered and are not triggered by the patient.
17.2.5.3 PC - SIMV
Pressure-controlled
Time-controlled
Machine-triggered
Synchronized inspiration
In PC - SIMV, the patient can breathe spontaneously at any time, while the number
of mandatory breaths is predefined. When synchronization is switched on, the
breaths are adapted to the spontaneous breathing efforts of the patient. If
spontaneous breathing effort by the patient is detected during the inspiratory trigger
window, a patient-triggered breath will be initiated.
17.2.5.4 PC - SIMV / PS
This mode is similar to PC - SIMV, except that the patient's spontaneous breathing
at the PEEP level during the expiratory phase is pressure-supported with Psupp
when outside the trigger window.
36259
7ULJJHUZLQGRZ
LQVS
7LPH
7LPH
0DFKLQHWULJJHUHGEUHDWK
3DWLHQWWULJJHUHGEUHDWK
Trigger
Flow
Slope
Slope
Paw
1/RR
Psupp
PEEP
Ti
Insp term
Pinsp
Trigger

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals