Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 305
Principles of operation
17.2.6.4 VC - SIMV
Volume-controlled
Pressure-limited
Time-controlled
Machine-triggered or patient-triggered
Constant inspiratory flow
Synchronized inspiration
In VC - SIMV, the patient can breathe spontaneously at any time, while the number
of mandatory breaths is predefined. When synchronization is switched on, the
breaths are adapted to the spontaneous breathing of the patient. If inspiratory effort
by the patient is detected during the inspiratory trigger window, a patient-triggered
breath will be initiated.
17.2.6.5 VC - SIMV / PS
This mode is similar to VC - SIMV, except that the patient's spontaneous breathing
at the PEEP level during the expiratory phase is pressure-supported with Psupp
when outside the trigger window.
17.2.6.6 VC - AF
With AutoFlow, the set tidal volume VT is applied for all mandatory volume-
controlled breaths with the lowest required pressure. The pressure patterns and
flow patterns of the mechanical inspiratory breaths correspond to those of pressure-
controlled ventilation.
36290
7ULJJHUZLQGRZ
LQVS
7LPH
7LPH
0DFKLQHWULJJHUHGEUHDWK
3DWLHQWWULJJHUHGEUHDWK
H[S
Flow
Paw
Pmax
PEEP
Ti
1/RR
Psupp
Insp term
Trigger
Slope
Slope
Trigger

Table of Contents

Other manuals for Dräger Atlan A350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 and is the answer not in the manual?

Dräger Atlan A350 Specifications

General IconGeneral
BrandDräger
ModelAtlan A350
CategoryMedical Equipment
LanguageEnglish

Related product manuals