EasyManuals Logo

Dräger Atlan A350 User Manual

Dräger Atlan A350
332 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
38 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Overview
3.2.12.2 Backup manual mode
In various technical fault situations, the backup manual mode enables a direct
changeover to manual ventilation in order to continue the therapy.
Backup manual mode with an electronically controlled gas mixer
1. Open the flap (1).
2. Activate the backup manual switch (2). Follow the instructions on the product
label (3) or the screen (4).
3. Set the O
2 switch to Add. O2.
4. Open the flow control valve on the emergency O
2 delivery and set an adequate
O
2 flow. The set O2 flow constitutes the total fresh-gas flow.
5. Ventilate the patient manually.
a. During mechanical ventilation or in the Man/Spon ventilation mode:
Ventilate manually with the breathing bag.
b. When using the external fresh-gas outlet:
Ventilate manually with the breathing bag on the non-rebreathing system.
6. Check the vaporizer setting.
4067639762